SHANGHAI MICROCRE OPTICS-MECH TECH CO., LTDAbout MICROCRE400-812-7833
SOLUTIONS

Quantitative microstructure, particles and defects

Turn images into auditable measurements by defining field selection, scale calibration, lighting, segmentation and statistical units together. Applications include layer thickness, particle size and 2D area fractions. Keep sectional results distinct from 3D volume, true morphology and defect incidence.

Theory & worked calculations

Item

Translation pending review.

二维孔隙面积:从像素到面积分数:模型与量纲

Equation

Ap=Np s²;Avalid=Nvalid s²

Variables & dimensions

Translation pending review.

有效像素数=1000000 px;孔隙像素数=24000 px;比例尺s=0.5 µm/px

Assumptions & limits

Translation pending review.

本例没有给出真实三维孔隙率。分割误差、制样拔出、孔隙盖层、视场选择及截面代表性均可主导最终偏差。

Worked example

Translation pending review.

假设输入演算:24000×0.5² = 6000 µm²;1000000×0.5² = 250000 µm²;6000/250000×100 = 2.4 %。同一图像统一缩放时,像素数比仍可保持2.4%,但绝对面积和粒径需要正确校准。这个区别解释了为什么面积分数正常并不能证明尺度标定正确。

Reporting requirements

Translation pending review.

以试样编号关联原始图像、测量序列、设备配置及方法版本。

Translation pending review.

分别报告重复性、异常处理及本模型未覆盖的影响因素,不用演算示例替代实测。

Common misconceptions

Translation pending review.

数字放大不能恢复光学分辨能力;比例尺、像元采样与边界识别应分别核查。

Translation pending review.

一个试样的多个视场不是多个独立试样;二维面积结果不自动等于三维体积指标。

Workflow & steps

01

Define metrics and sampling

Specify equivalent diameter, length, thickness or area fraction and define border-object and touching-object rules. Sample multiple fields across the specimen rather than selecting only clear or typical areas; retain field coordinates.

02

Prepare and calibrate imaging

Match target scale to optical resolving capability and calibrate the actual objective–camera setup. Fix illumination, exposure and white balance and retain minimally compressed originals. Recheck scale after optical or acquisition-size changes.

03

Create interpretable segmentation

Set thresholds or classification rules on representative images and retain mask overlays. Review low contrast, scratches, pores and phase boundaries manually. Validate on independent images before batch processing instead of tuning toward an expected conclusion.

04

Summarise and check sensitivity

Compute distributions and variability at the predefined object and field levels; test sensitivity to small threshold changes. Divide target area by valid field area for area fraction, explicitly defining masks, holes and edge rules.

05

Deliver reproducible data

Report specimen count, fields, analysed area, algorithm and software version, with originals, masks, scale and result tables. Identically named metrics may differ across software; compare definitions before comparing values.

Checks

  • • Fix scale and acquisition configuration
  • • Retain masks and review records
  • • State the interpretation limits of 2D results

Illustrative example

Teaching example: apply two reasonable thresholds to the same ceramic sections and compare sensitivity of 2D pore area fraction. This illustrates processing effects and is not a measured volumetric-porosity report.

Related instruments & resources

Links support navigation and do not establish a model’s performance or standards compliance.

Your next research question. Let’s explore.

Tell us about your material, specimen and measurement goal.

Discuss your application ↗
Find your instrument
⌕