SHANGHAI MICROCRE OPTICS-MECH TECH CO., LTDAbout MICROCRE400-812-7833
METHODS

High-temperature indentation: aligning temperature, load and displacement

High-temperature indentation couples local deformation measurements to a thermal environment. Key issues include contact temperature, thermal drift, tip stability and time-dependent material deformation. Distinguish optical measurement of residual indents from instrumented force–displacement testing; their analysis methods and applicability differ.

Theory & worked calculations

Item

Translation pending review.

高温压痕:漂移预算与蠕变分离:模型与量纲

Equation

δdrift=v×t;δremaining=δtotal−v×t

Variables & dimensions

Translation pending review.

总位移=600 nm;名义漂移速率=2 nm/s;时间=60 s;漂移下界=1.5 nm/s;漂移上界=2.5 nm/s

Assumptions & limits

Translation pending review.

线性修正只在漂移近似恒定且独立可辨识时成立。非线性热过渡、压头反应、蠕变和接触重排会破坏这个简化;不要把剩余位移直接命名为材料本征蠕变。

Worked example

Translation pending review.

假设输入演算:2×60 = 120 nm;600−2×60 = 480 nm;600−2.5×60 = 450 nm;600−1.5×60 = 510 nm。名义漂移占总位移20%。只看600 nm曲线会把仪器影响和材料响应混在一起;即使相减后得到480 nm,也需证明接触面积、加载状态和温度稳定后才能进一步建立蠕变模型。

Reporting requirements

Translation pending review.

以试样编号关联原始图像、测量序列、设备配置及方法版本。

Translation pending review.

分别报告重复性、异常处理及本模型未覆盖的影响因素,不用演算示例替代实测。

Common misconceptions

Translation pending review.

炉温稳定不等于接触区温度准确或测量链无漂移;需独立定位温度与位移基准。

Translation pending review.

保持段位移可能同时包含真实变形与仪器漂移,不能把全部斜率直接扣除。

Workflow & steps

01

Define the measurand

Specify whether the goal is hardness, indentation modulus, creep response or crack observation. Record specimen, atmosphere and surface state. Select tips and fixtures compatible with both temperature and material; a single temperature limit does not define the complete configuration's suitability.

02

Document the thermal state

Document sensor locations, calibration and heating history. Test after the specimen, tip vicinity and load path approach stable conditions. Record specimen-surface temperature separately from the furnace setpoint and recheck stability after rapid specimen exchange.

03

Assess drift and baseline

Record displacement baseline and drift rate under defined loading and contact conditions, and inspect the tip before and after testing. Continuing displacement may contain both thermal drift and real creep; the entire hold-segment slope must not automatically be subtracted as drift.

04

Acquire and analyse synchronously

Synchronously retain force, displacement, time, temperature and images. Instrumented hardness may be calculated as H=Pmax/Ac; Pmax in mN and Ac in μm² give H in GPa. However, validate the contact-area function, frame compliance and pile-up or sink-in effects. Maximum depth is not a substitute for contact depth.

05

Repeat and report limitations

Use repeated points, reference materials or measurements after cooling to investigate changes. Report atmosphere, thermal hold, drift treatment and anomalous curves. Account for oxidation, phase changes and tip reactions when interpreting elevated-temperature results.

Checks

  • • State sensor locations and uncertainty sources for contact temperature.
  • • Validate each atmosphere, tip and material combination separately.
  • • Keep raw and corrected curves, stating the correction model and its limits.

Illustrative example

Teaching example: independently estimated steady drift of 2 nm/s corresponds to 120 nm over 60 s. Use this in an error budget; a linear correction is meaningful only when drift is approximately constant and distinguishable from material deformation.

Related instruments & resources

Links support navigation and do not establish a model’s performance or standards compliance.

Your next research question. Let’s explore.

Tell us about your material, specimen and measurement goal.

Discuss your application ↗
Find your instrument
⌕