SHANGHAI MICROCRE OPTICS-MECH TECH CO., LTDAbout MICROCRE400-812-7833
INDUSTRIES

Semiconductors, packaging and microelectronics

Create a path from locating small targets to section analysis. For package interfaces, joints, layers and local defects, control material removal, edge retention and imaging capability. Optical microscopy provides sectional morphology and dimensions; finer structures need other suitable techniques.

Theory & worked calculations

Item

Translation pending review.

封装薄层:五点厚度与离散性:模型与量纲

Equation

Translation pending review.

t平均=s×mean(N);样本标准差

Variables & dimensions

Translation pending review.

像素厚度=[41, 43, 40, 44, 42] px;s=0.2 µm/px;位置数=5 个

Assumptions & limits

Translation pending review.

测量线彼此可能相关,不能直接将所有位置当作独立试样。二维截面不足以证明电学功能或界面结合强度。

Worked example

Translation pending review.

假设输入演算:0.2×(41+43+40+44+42)/5 = 8.4 µm;sqrt[Σ(ti−8.4)²/(5−1)] = 0.316228 µm。0.316 µm描述这五个位置的离散性,尚未合入比例尺校准和边缘识别误差。即使位置数增加,校准系统误差也不会按1/√n消失。

Reporting requirements

Translation pending review.

以试样编号关联原始图像、测量序列、设备配置及方法版本。

Translation pending review.

分别报告重复性、异常处理及本模型未覆盖的影响因素,不用演算示例替代实测。

Common misconceptions

Translation pending review.

数字放大不能恢复光学分辨能力;比例尺、像元采样与边界识别应分别核查。

Translation pending review.

一个试样的多个视场不是多个独立试样;二维面积结果不自动等于三维体积指标。

Workflow & steps

01

Locate the target and preserve condition

Record orientation, target position and existing defects; define interfaces and dimensions. For buried targets, use available localisation information to plan the approach and avoid grinding past the area of interest.

02

Support and remove material precisely

Choose preparation compatible with metals, ceramics and polymers; avoid unsuitable mounting temperatures for sensitive parts. Use controlled cutting and progressive target approach, inspecting position and recording remaining distance at each stage.

03

Control interfacial relief

Hardness differences can produce relief, smearing or embedded abrasive; control media and process conditions. Before calling cracks or voids manufacturing defects, examine preparation-induced cracking, pull-out and resin gaps.

04

Calibrate and quantify

Calibrate the actual objective and camera, standardise exposure and measure thickness, spacing and defects. Compare sections or fields and state directions and boundary rules. Sectional dimensions are not complete 3D defect dimensions.

05

Assemble failure-analysis evidence

Archive original images, preparation path, location maps and measurements together; distinguish observations from mechanism hypotheses. Electrical failure or chemical contamination needs corresponding tests, not a final conclusion from one section image.

Checks

  • • Record target position and removal path
  • • Check thermal sensitivity and mixed hardness
  • • Separate observed morphology from failure inference

Illustrative example

Teaching example: compare void morphology in successive package-joint sections and check polishing smear. This illustrates reasoning and is not a claimed case from a semiconductor customer.

Related instruments & resources

Links support navigation and do not establish a model’s performance or standards compliance.

Your next research question. Let’s explore.

Tell us about your material, specimen and measurement goal.

Discuss your application ↗
Find your instrument
⌕
Search products, cases, solutions, standards and literature.