Novel high temperature vacuum nanoindentation system with active surface referencing and non-contact heating for measurements up to 800 °C

Translation pending review.

通过独立参考轴、非接触加热及真空环境,研究高温位移测量的稳定性。

2019Review of Scientific Instrumentsabstractabstract2026-11-16Marcello Conte, Gaurav Mohanty, Jakob J. Schwiedrzik, Jeffrey M. Wheeler, Bertrand Bellaton, Johann Michler, Nicholas X. Randall

Original paper title retained. Summaries are editorial notes; obtain full text from the publisher or lawful open sources.

Findings

Finding

Translation pending review.

论文报告的最低热漂移小于2 nm/min。

Conditions

Translation pending review.

UNHT3 HTV系统;铜、熔融石英和硬质涂层;最高800°C、真空约10⁻⁷ mbar的研究条件。

Source location

Translation pending review.

摘要

Quantity
Value

2

Unit

Translation pending review.

nm/min(小于)

Application relevance

Translation pending review.

Translation pending review.

用于理解参考测量、加热与环境控制对高温压痕的共同作用。

Limitations

Translation pending review.

Translation pending review.

该数值属于文中外部系统的实验结果;最低值不等于所有温度、材料上的保证值。