Spectral verification of cast aluminium incoming lots
Each aluminium ingot lot needs major and trace element checks for release, with one operator and under three minutes per piece.
The constraint
Segregation scatters single-spot results, and oxide or oil on the surface destabilises excitation.
Approach
Prepare the excitation surface to standard, take three spots with two burns each, judge on the median and re-test if out of limit.
- Turn or mill the surface to remove oxide and oil
- Load the grade curve, three spots, two burns each
- Judge on the median, record out-of-limit items
Outcome
Cycle time 2.5 min, agreeing with third-party checks; segregated lots now use five spots.
Related instruments
Application scenario files: industry, workpiece type and measurement goals come from routine engineering discussions. Customer names and specific data are anonymized; nothing here is a traceable commercial record or a commitment about any customer's production results.