APPLICATION FILE / Precision optics and metrology

Calibrating the indexing error of a precision rotary table

A rotary table used for optical assembly must be checked every 15° to confirm angular margin for mirror-group alignment.

Industry:Precision optics and metrologyInstruments:MC030-1401 · 1401-15

The constraint

The encoder readout cannot verify itself, and polygon face errors leak into autocollimator results.

Approach

Use a polygon as the angle reference, measure in both directions and separate polygon from table error by closure.

  1. Level the table and align the autocollimator axis
  2. Read at every 15° forward and reverse
  3. Correct by closure and issue an indexing table

Outcome

Peak indexing error 18″ against a 30″ allowance; the table ships with the stage and is re-checked every six months.